Durom™ Hip Resurfacing

Component Sizing

Durom Acetabular Component Durom Femoral Component
Size Inner
Diameter
Code Size Inner
Diameter
Code
44 38mm D 38 38mm D
46 40mm F 40 40mm F
48 42mm H 42 42mm H
50 44mm J 44 44mm J
52 46mm L 46 46mm L
54 48mm N 48 48mm N
56 40mm P 50 50mm P
58 52mm S 52 52mm S
60 54mm T 54 54mm T
62 56mm V 56 56mm V
64 58mm X 58 58mm X
66 60mm Z 60 60mm Z

  • 12 different sizes are offered in a size increment of 2mm. The wide range of sizes offers an optimal fit.
  • The acetabular component is designed to be implanted without cement. The femoral component is designed to be inserted with cement.
  • The actual diameter of the acetabular component is 2 mm greater than its nominal diameter, e. g. last used reamer -- 54 mm, trial implant 54 mm (the trial implant is line to line with the reamer), acetabular implant size 54/N. This will result in a 2 mm hemispherical press-fit.
  • Large diameter modular Durom Metasul femoral heads for use with the range of Centerpulse stems are available and must be used in conjunction with the Durom acetabular components.